A compact and high-precision fluid warming system designed to rapidly warm blood and IV fluids, helping prevent hypothermia during transfusion and infusion procedures. Its dual-channel design ensures efficient and consistent temperature control for critical care environments.
• Blood Transfusion
• High Volume Infusion Therapy
• ICU & Critical Care
• Operation Theatres
• Emergency & Trauma Care
• Neonatal & Pediatric Care
A light, ergonomic design and convenient trolley mounting options make the Blanket Warming Unit a desirable choice for patient warming needs.
Dual Channel Heating
Simultaneously warms fluids through two independent heating channels for faster and efficient performance.
Precision Temperature Control
Adjustable temperature range from 32°C to 43°C with fine control for safe and accurate warming.
Rapid Heating System
Quickly raises fluid temperature to desired levels, reducing patient heat loss.
Lightweight & Portable Design
Lightweight and portable design for easy manoeuvrability.
Intelligent Display Interface
Shows real-time temperature, set values, working time, and system status.
Advanced Safety Protection
Built-in high & low temperature alarms, overcurrent protection, and automatic shutdown for patient safety.
Compact & Portable
Lightweight (~1 kg) design makes it easy to use in ICU, OT, and emergency setups.
| Parameter | Specification |
|---|---|
| Model | DEXUP FW D 100 |
| Design | Control device with 2 heating silicon tubes |
| Temperature Setting Range | 32°C-43°C (0.1°C step) |
| Liquid Outlet Temperature Deviation | ±2.0°C |
| Heating Time | From 23 ± 2°C to 36°C in a short time |
| Display Parameters | Current temperature, Set temperature, Running/Working time, Channels, Alarm mute, Seek |
| Safety - High Temperature Protection | Alarm and stop if temperature reaches 45°C for 30s or exceeds set temperaturo by> 8°C |
| Safety - Low Temperature Protection | After 5 minutes from startup, if tube temperature $30°C, alarm and stop |
| Overcurrent Protection | Dual input fuse circuit (Fuse: 5T 4A/250V) |
| Fault Indication | Display shows fault codes and fault information |
| Power Supply | 220 VAC, 50 Hz, 4 A |
| Power Cord | 5.0 m, 250 V, 10 A |
| Display Size | 4.3 inches |
| Weight | Approx. 1 kg |
| Dimensions (cm) | 15 (W) x 7.5 (D) x 20 (H) |
| Service Life | 5 years |
| Operating Environment | Temperature: 5-30°C; Humidity: 30-75% RH; Pressure: 700-1060hPa |
| Standard Accessories |
Fluid Warmer x1; Heating Silicon Tube x2; Detachable Power Cord x1; User Manual x1 |
| Application | Immediate heating of blood transfusion or infusion fluids: reduces temperature difference between liquid and patient's body to prevent hypothermia |